ISCEV standard pattern reversal VEP development: paediatric reference limits from 649 healthy subjects.
Dorothy A Thompson1,2, Eszter Mikó-Baráth3, Sharon E Hardy4
1The Tony Kriss Visual Electrophysiology Unit, Clinical and Academic, Department of Ophthalmology, Great Ormond Street Hospital for Children NHS Trust, London, UK. dorothy.thompson@ucl.ac.uk.
Documenta Ophthalmologica. Advances in Ophthalmology
|November 8, 2023
Summary
This study established reference intervals for pattern reversal visual evoked potentials (prVEPs) in children. Pooled data from three centers showed comparable results, with no clinically significant sex differences, providing valuable prVEP normative data.
Area of Science:
- Ophthalmology
- Neuroscience
- Pediatrics
Background:
- Visual evoked potentials (VEPs) are crucial for assessing visual pathway function.
- Standardized protocols, like the ISCEV standard for pattern reversal VEPs (prVEPs), are essential for reliable clinical interpretation.
- Establishing normative data across different age groups and centers is vital for accurate diagnosis.
Purpose of the Study:
- To determine the agreement of prVEPs across three European centers.
- To identify any influence of sex on prVEP measurements.
- To establish reference intervals for prVEPs from birth to adolescence.
Main Methods:
- PrVEPs were recorded from healthy infants and children (2 weeks to 16 years) across three centers.
- Data were analyzed using nonlinear quantile Bayesian regression to model amplitudes and peak times.
- Comparisons were made between centers and sexes, and a 'critical age' for P100 peak time stabilization was determined.
Main Results:
- Data from the three centers showed adequate comparability, with no clinically meaningful sex differences.
- P100 peak time stabilized by 27 and 34 weeks for large and small check sizes, respectively.
- Post-critical-age reference limits for P100 peak time were 87-115 ms (large checks) and 96-131 ms (small checks); amplitude ranges were 5-57 μV and 3.5-56 μV.
Conclusions:
- PrVEP reference data are combinable across centers within ISCEV Standard tolerances, highlighting the standard's clinical utility.
- The established reference data can be validated or adapted for use in other settings.