Related Experiment Video
Updated: Jul 20, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Computed Tomography-Based Humeral Templating for Uncemented Elbow Arthroplasty
Morgan R Kohls1, Emerald D Robertson1, Colin H Beckwitt1
1Department of Orthopaedic Surgery, University of Pittsburgh Medical Center, PA, USA.
Background:
Intramedullary (IM) screw insertion into the distal humerus provides fixation for a novel, uncemented elbow arthroplasty. A multitude of screw sizes is required to accommodate variable humeral morphology. The goal of this study was to use computed tomography (CT) for IM screw sizing and to validate this templating by inserting screws into three-dimensionally (3D) printed models.
Methods:
Computed tomography humerus scans for 30 patients were reformatted in the plane of the distal IM canal. Screw size was templated by measuring the canal diameter at 3 locations corresponding to the lengths of the screws being tested. Interrater and intrarater reliabilities of the measurements were assessed. Three-dimensional models of 5 humeri were printed, and IM screws were placed to achieve a secure endosteal fit.
Results:
We identified combinations of body components and IM screw length and diameter for all patients to seat this uncemented elbow arthroplasty. The measurements and screw width determinations were reliable. Canal diameter correlated with age but was unrelated to sex. Screws were inserted into five 3D-printed models which matched the templates and demonstrated mechanical and radiographic evidence of secure fit.
Conclusions:
This study characterizes distal humerus anatomy in the context of IM screw fixation. Humerus CT scans of 30 patients were able to be templated, and validation via implantation of IM screws into 3D models was successful. Computed tomography templating will allow surgeons to predict the optimal screw size prior to implantation. A broad range of screw lengths and diameters is critical for implantation of this novel elbow arthroplasty.
Related Concept Videos
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Bones of the Upper Limb: Humerus