Related Experiment Video
Updated: Jul 11, 2025

Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
A bending model for assessing relative stiffness and strength of orthopaedic fixation constructs
Bijan Abar1, Elijah Vail2, Elizabeth Mathey3
1Duke University, Department of Mechanical Engineering and Material Sciences, USA; Duke University, Department of Orthopaedic Surgery, USA.
Background:
The purpose of this study is to develop a simple and reproducible bending model that is compatible with a wide range of orthopaedic fixation devices and 3D printed spacers.
Methods:
A robust 4-point bending model was constructed by securing sawbones blocks with different orthopaedic fixation device constructs. Stress strain curves derived from a fundamental mechanics model were used to assess the effect of bone density, type of hardware (staple vs intramedullary beam), the use of dynamic compression, orientation of staples (dorsal vs plantar), and the use of 3D printed titanium spacers.
Findings:
The high throughput 4-point bending model is simple enough that the methods can be easily repeated to assess a wide range of fixation methods, while complex enough to provide clinically relevant information.
Interpretations:
It is recommended that this model is used to assess a large initial set of fixation methods in direct and straightforward comparisons.
Related Concept Videos
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each...
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Temperature Dependent Deformation
Flexural Stress
Hooke's Law states that within the material's elastic limits, stress is directly proportional to strain. In a member experiencing a bending moment, the strain at any point is relative to...
Plastic Deformations
Hooke's Law