Feline Heinz bodies interfering in standard bicarbonate analysis by the ABL90 FLEX
Emma Strage1,2, Anna Hillström1,2, Ulrika Falkenö2
1Department of Clinical Sciences, Swedish University of Agricultural Sciences, Uppsala, Sweden.
Abstract:
Venous blood gases were analyzed with ABL90 FLEX in two cats with Heinz bodies in approximately 60% of the erythrocytes. The instrument demonstrated an inability to correctly report standard bicarbonate (stHCO3 - ), hematocrits, and hemoglobin concentrations by indicating an OXI spectrum mismatch alarm (ie, the spectrum of measured hemoglobin forms differed from the spectrum of calculated forms). Actual bicarbonate (aHCO3 - ) did not indicate any errors. The ABL90 FLEX uses spectrophotometry to measure hemoglobin, and the presence of Heinz bodies interfered with the measurement in these cases. Because hemoglobin is included in the formula for calculating stHCO3 - , the instrument gave an alarm for stHCO3 - . At follow-up, Heinz bodies were present in only 2%-3% of the erythrocytes, and the ABL90 FLEX did not indicate any alarm messages. To the authors' knowledge, these are the first cases reported that have interference in stHCO3 - measurements due to Heinz body formation using the ABL90 FLEX, a common blood gas instrument used in both veterinary and human critical care. The methodology used for evaluating acid-base status should be taken into consideration, and caution is needed when interpreting acid-base results in cats with Heinz bodies.
More Related Videos
15:04PeptiQuick, a One-Step Incorporation of Membrane Proteins into Biotinylated Peptidiscs for Streamlined Protein Binding Assays
Published on: November 2, 2019
08:42Using Multi-fluorinated Bile Acids and In Vivo Magnetic Resonance Imaging to Measure Bile Acid Transport
Published on: November 27, 2016
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...