Measuring Carbon-based Contaminant Mineralization Using Combined CO2 Flux and Radiocarbon Analyses
Published on: October 21, 2016
Stable isotope analysis of atmospheric CO2 using a Gasbench II-Cold Trap-IRMS setting
Simon Leitner1, Kathiravan Meeran1, Andrea Watzinger1
1Institute of Soil Research, University of Natural Resources and Life Sciences, Vienna, Tulln, Austria.
Rationale:
The measurement of the stable carbon and oxygen isotope ratio of (atmospheric) carbon dioxide (CO2 ) is a useful technique for the investigation and identification of the sources and sinks of the most abundant greenhouse gases by far. For this reason, we are presenting a measuring system here that enables a wide range of users to carry out stable isotope analysis of atmospheric CO2 using off-the-bench hardware and software.
Methods:
The fully automated system uses cryogenic and gas chromatographic separation to analyse CO2 from 12-mL whole air samples and consists of an autosampler, a Gasbench II (GB), a downstream cryo trap and a continuous flow gas interface feeding into a sector field mass spectrometer (GC Pal/GB/Cold Trap/ConFlo IV/DeltaV Plus). The evaluation of the system performance was based on the analysis of samples prepared from eight CO2 sources (four CO2 reference gases and four artificial air tanks).
Results:
The overall measurement uncertainty (averaged single standard deviation (1σ) of measurement replicates from each CO2 source) in the determination of the carbon and oxygen isotope ratio was 0.04‰ and 0.09‰ (n = 24). Furthermore, we were able to show that the measurement data also allowed for the quantification of the CO2 mole fraction, with a precision of 1.2 μmol mol-1 in the analysis range of 400-500 μmol mol-1 .
Conclusions:
Our protocol provides a detailed description of the measurement set-up and the analysis procedure, how raw data should be evaluated and gives recommendations for sample preparation and sampling to enable a fully automated whole air sample analysis. The quantification limit of CO2 mole fractions and measurement precision for carbon and oxygen isotope ratios of CO2 should meet the requirements of a wide range of users.