Automated, High-resolution Mobile Collection System for the Nitrogen Isotopic Analysis of NOx
Published on: December 20, 2016
Establishment of a Purge and Trap Continuous-Flow Isotope Ratio Mass Spectrometer System for Analysis of Stable
Monis Nolitha Gcakasi1, Christine Stumpp1, Andrea Watzinger2
1Institute of Soil Physics and Rural Water Management, Department of Landscape, Water and Infrastructure, BOKU University, Vienna, Austria.
Rationale:
Pollution of surface and shallow groundwater by nitrate ( ) is a global concern, resulting in the deterioration of drinking water quality. Stable isotopes of (δ15N and δ18O) can be used to trace its sources and identify prevailing biogeochemical processes.
Methods:
Conversion of aqueous to N2O headspace gas by Ti(III) reduction is a new method for analysis of stable isotopes. Previous literature introduces the analytical procedure but provides limited guidelines for instrument set-up and operation. Here, we present an automated purge-and-trap isotope ratio mass spectrometer (P&T-IRMS) combined with the Ti(III) reduction method for analysis of δ15N and δ18O in .
Results:
The P&T-IRMS base analytical precision was ±0.3‰ and ±0.2‰ for δ15N and δ18O, respectively. The limit of quantification (LOQ) for the N2O gas standard was 1 μL L-1 (1.1 nmol N2O-N) for both δ15N and δ18O. The range for accurate measurement was 0.2-0.3 mg L-1 (3.3-6.6 nmol). Based on the logarithmic trend of the N2O peak area vs. R15N/14N and R18O/16O, δ15N and δ18O corrections are presented for values outside the range of accuracy. A comparison of δ15N and δ18O from P&T-IRMS and Ti(III) reduction measurements with EA-IRMS values showed high accuracy. The measurement precision and uncertainties for our internal standard were ±0.2 (±0.6) and ±0.2 (±0.9) for δ15N and δ18O, respectively.
Conclusion:
The P&T-IRMS and Ti(III) reduction method demonstrated acceptable accuracy and precision, similar to well-established methods for analysis of stable isotopes. This publication will assist laboratories which utilize IRMS headspace gas instrumentation with the process of IRMS set-up and operation, establishment of an independent analytical procedure for the Ti(III) reduction method, and data evaluation and correction.