Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Relation of Crown Failure Load to Flexural Strength for Three Contemporary Dental Polymers
1Department of Substitutive Dental Sciences, Taibah University, Madinah 42353, Saudi Arabia.
Abstract:
Polymeric materials show great promise for use in a variety of dental applications. Manufacturers generally provide flexural strength information based on standardized (ISO and ASTM) specimen dimensions and loading conditions. It is not clear, however, if flexural strength data are predictive of the clinical performance of dental crowns. The objectives of this study were, therefore, to determine whether flexural strengths, as measured via three-point bending (3PB), would be predictive of failure loads assessed via crunch-the-crown (CTC) tests. Three brands of polymers (Trilor, Juvora, and Pekkton) were fabricated into rectangular bars and fully contoured crowns (10 specimens of each polymer brand, 30 specimens of each shape). Differential scanning calorimetry (DSC), Fourier transform infrared spectroscopy (FTIR), and burn off tests were used to characterize/confirm the materials. Bars were tested blindly in 3PB to determine flexural strength, and crowns were CTC-tested to determine failure load after luting to resin abutments. The statistical significance of the test results was evaluated via one-way ANOVA (α = 0.05) and Pearson's correlation coefficient, while regression analysis was used to test for a correlation between 3PB and CTC results. The fracture mechanisms and failure surface characteristics were characterized using scanning electron microscopy (SEM). There were significant differences (p < 0.05) in the mean crown failure loads (Trilor (7033 N) > Juvora (5217 N) > Pekkton (3023 N)) and mean flexural strengths of the bars (Trilor (468 MPa) > Juvora (197 MPa) = Pekkton (192 MPa)). The mode of crown fracture was different between the materials and included deformation (Juvora), ductile-to-brittle fracture (Pekkton), and a combination of cracks and deformation (Trilor). Flexural strengths did not correlate with the corresponding crown failure loads for any of the materials tested. These results suggest that dental practitioners should not rely on the flexural strengths reported from three-point bending tests, as advertised by the manufacturer, to predict the performance of polymeric crowns.
More Related Videos
Related Concept Videos
Members Made of Elastoplastic Material
As the bending moment...
Plastic Deformations
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...
Flexural Stress
Hooke's Law states that within the material's elastic limits, stress is directly proportional to strain. In a member experiencing a bending moment, the strain at any point is relative to...
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each...
Plastic Behavior