WATCH-BPM-Comparison of a WATCH-Type Blood Pressure Monitor with a Conventional Ambulatory Blood Pressure Monitor and
Mathini Vaseekaran1, Sven Kaese1, Dennis Görlich2
1Department of Cardiology and Critical Care Medicine, Johannes Wesling University Hospital, 32429 Minden, Germany.
Background:
Smart devices that are able to measure blood pressure (BP) are valuable for hypertension or heart failure management using digital technology. Data regarding their diagnostic accuracy in comparison to standard noninvasive measurement in accordance to Riva-Rocci are sparse. This study compared a wearable watch-type oscillometric BP monitor (Omron HeartGuide), a wearable watch-type infrared BP monitor (Smart Wear), a conventional ambulatory BP monitor, and auscultatory sphygmomanometry.
Methods:
Therefore, 159 consecutive patients (84 male, 75 female, mean age 64.33 ± 16.14 years) performed observed single measurements with the smart device compared to auscultatory sphygmomanometry (n = 109) or multiple measurements during 24 h compared to a conventional ambulatory BP monitor on the upper arm (n = 50). The two BP monitoring devices were simultaneously worn on the same arm throughout the monitoring period. In a subgroup of 50 patients, single measurements were also performed with an additional infrared smart device.
Results:
The intraclass correlation coefficient (ICC) between the difference and the mean of the oscillometric Omron HeartGuide and the conventional method for the single measurement was calculated for both systole (0.765) and diastole (0.732). This is exactly how the ICC was calculated for the individual mean values calculated over the 24 h long-term measurement of the individual patients for both systole (0.880) and diastole (0.829). The ICC between the infrared device and the conventional method was "bad" for SBP (0.329) and DBP (0.025). Therefore, no further long-term measurements were performed with the infrared device.
Conclusion:
The Omron HeartGuide device provided comparable BP values to the standard devices for single and long-term measurements. The infrared smart device failed to acquire valid measurement data.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: