Robust Fluorescence Collection Module for Wide-Bore Ion Cyclotron Resonance Mass Spectrometers
Francis O Talbot1, Cynthia M Suarez1, Andrea M Nagy1
1Department of Chemistry, University of Toronto, Toronto, Ontario M5S 3H6, Canada.
Abstract:
Mass spectrometers are at the heart of the most powerful toolboxes available to scientists when studying molecular structure, conformation, and dynamics in controlled molecular environments. Improved molecular characterization brought about by the implementation of new orthogonal methods into mass spectrometry-enabled analyses opens deeper insight into the complex interplay of forces that underlie chemistry. Here, we detail how one can add fluorescence detection to commercial ultrahigh-resolution Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometers without adverse effects to its preexisting analytical tools. This advance enables measurements based on fluorescence detection, such as Förster resonance energy transfer (FRET), to be used in conjunction with other MS/MS techniques to probe the conformation and dynamics of large biomolecules, such as proteins and their complexes, in the highly controlled environment of a Penning trap.
More Related Videos
Related Concept Videos
Mass Analyzers: Common Types
Mass Spectrometers
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview