Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to
Thamonwan Chirdchid1, Sirinya Ruangchan2,3, Taweap Sanghangthum1,2,4
1Department of Radiology, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
Purpose:
The focus of this work was given on the relative stopping power (RSP) using the water equivalent thickness (WET) validation on tissue substitutes and real pig organs, as well as a dosimetric comparison of proton treatment plans between single-energy computed tomography (SECT) and dual-energy computed tomography (DECT)-based dose calculations.
Materials And Methods:
The CT calibration curve of SECT and DECT data was generated using the stoichiometric calibration method. WET measurement was performed for RSP validation using a Giraffe dosimeter (IBA dosimetry) in various substitute tissues (Gammex) and real pig tissues. The thorax (008A, CIRS) and head (731-HN, CIRS) phantoms were used to generate proton plans. The dosimetric evaluations of SECT and DECT-based plans were performed using the gamma analysis with 1%/1 mm and the dose-volume histograms (DVHs) comparison.
Results:
For RSP validation of substitute tissues, the largest percent WET difference between measurement and calculation was observed up to 17.9% (4 mm) in lung tissue, using SECT based. In real pig tissues, the average WET difference was 2.3% ± 2.1% and 2.5% ± 2.3% for SECT and DECT, respectively. The average gamma passed of about 92.1% for the lung and 96.8% for the head regions was reported. For the lung region, the DVH of the target dose was observed with a higher predicted dose in SECT than in DECT, while results in the head region were in good agreement for both SECT and DECT.
Conclusion:
The performed dosimetric comparison indicates the dose differences between SECT and DECT. The impact of the CT calibration curve is more pronounced for the thorax region.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Biological Effects of Radiation
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET