A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
Published on: February 7, 2015
Novel Peripheral Perfusion Dynamics Indices for Detecting and Grading Arterial Stenosis
Tomer Joshua Heitner1, Amit Livneh1, Amir Landesberg2
1Faculty of Biomedical Engineering, Technion-Israel Institute of Technology, Technion City, 32000, Haifa, Israel.
Purpose:
Peripheral artery disease causes severe morbidity, especially in diabetics and the elderly. There is a need for accurate noninvasive detection of peripheral arterial stenosis. The study has tested the hypothesis that arterial stenosis and the associated adaptation of the downstream circulation yield characteristic changes in the leg perfusion dynamics that enable early diagnosis, utilizing impedance plethysmography.
Methods:
The arterial perfusion dynamic was derived from impedance plethysmography (IPG). Two degrees of arterial stenosis were emulated by inflating a blood-pressure cuff around the thigh to 45 and 90 mmHg, in healthy volunteers (n = 30). IPG signals were acquired continuously throughout the experiment. Ankle and brachial blood pressures were measured at the beginning of each experiment and at the end of each emulated stenosis phase.
Results:
Thigh compressions did not affect the pulse-transit time, but prolonged the time to the peak perfusion wave. Segmentation of the perfusion upstroke into two phases, at the time point of maximum acceleration (MAT), revealed that arterial compression prolonged only the initial slow phase duration (SPd). The MAT and SPd were proportional to the emulated stenosis severity and detected the arterial stenosis with high sensitivity (> 93%) and specificity (100%). The SPd increased from 46.4 ± 21.2 ms at baseline to 75.4 ± 38.5 ms and 145 ± 39 ms under 45 mmHg and 90 mmHg compressions (p < 0.001), without affecting the pulse-transit time.
Conclusions:
The novel method and indices can identify and grade the emulated arterial stenosis with high accuracy and may assist in differentiating between focal arterial stenosis and widespread arterial hardening.
More Related Videos
10:27Real-Time Assessment of Spinal Cord Microperfusion in a Porcine Model of Ischemia/Reperfusion
Published on: December 10, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Pulse
The pulse serves as a clinical...
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...
Cardiovascular System Abnormal Findings I: Inspection and Palpation
Abnormal findings observed during an inspection