Background:
Valve frame infolding (VFI) is a rare complication (1-3%) of transcatheter aortic valve implantation (TAVI), which primarily occurs in large sized self-expanding valves. We report the incidence of VFI in a small group of patients with highly calcified and extra-large aortic annuli, who underwent implantation of the newer generation 34mm Medtronic Evolut Pro+ (EP+) valve system.
Methods:
A retrospective review was conducted on all patients presenting to a single centre experienced with TAVI in Sydney, NSW, Australia, between June and October 2022, for transfemoral TAVI using the Medtronic 34mm EP+ valve system. VFI was diagnosed using fluoroscopy, and pre-procedure computed tomography (CT) was analysed offline.
Results:
VFI occurred in four of 10 patients who underwent TAVI using the 34mm EP+ system. Between VFI and non-VFI patients, the annular size, aortic angulation and total calcium volume was similar. Calcium distribution between the coronary cusps was symmetrical in non-VFI patients (37%, 33% and 30%), compared to 52% in the non-coronary cusp (NCC) in VFI patients. The mean ellipticity index was higher in VFI versus non-VFI patients (22% vs 14%).
Conclusion:
The 34mm Medtronic EP+ valve system is vital for safe and effective treatment of a niche subgroup of patients with extra-large annuli and extensive calcification, however operators should be aware of the increased incidence of VFI. Heavy eccentric calcification at the NCC, as well as using the cusp-overlap implant technique may exacerbate VFI, while multiplanar fluoroscopic screening may improve recognition.
More Related Videos
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...