Accurate and continuous respiratory rate using touchless monitoring technology
Paul S Addison1, Carolyn Cohen2, Ulf R Borg3
1Medtronic Patient Monitoring, Edinburgh, UK.
Purpose:
Respiratory rate is a commonly used vital sign with various clinical applications. It serves as a crucial marker of acute health issues and any significant alteration in respiratory rate may be an early warning sign of major issues such as infections in the respiratory tract, respiratory failure, or cardiac arrest. Timely recognition of changes in respiratory rate enables prompt medical action, while neglecting to detect a change may lead to adverse patient outcomes. Here, we report on the performance of respiratory rate determined using a depth sensing camera system (RRdepth) which allows for continuous, non-contact 'touchless' monitoring of this important vital sign.
Methods:
Thirty adult volunteers undertook a range of set breathing rates to cover a target breathing range of 4-40 breaths/min. Depth information was acquired from the torso region of the subjects using an Intel D415 RealSense camera positioned above the bed. The depth information was processed to generate a respiratory signal from which RRdepth was calculated. This was compared to a manually scored capnograph reference (RRcap).
Results:
An overall RMSD accuracy of 0.77 breaths/min was achieved across the target respiratory rate range with a corresponding bias of 0.05 breaths/min. This corresponded to a line of best fit given by RRdepth = 1.01 x RRcap - 0.22 breaths/min with an associated high degree of correlation (R = 0.997). A breakdown of the performance with respect to sub-ranges corresponding to respiratory rates or ≤7, >7-10, >10-20, >20-30, >30 breaths/min all exhibited RMSD accuracies of less than 1.00 breaths/min. We also had the opportunity to test the performance of spontaneous breathing of the subjects which occurred during the study and found an overall RMSD accuracy of 1.20 breaths/min with corresponding accuracies ≤1.30 breaths/min across each of the individual sub-ranges.
Conclusions:
We have conducted an investigative study of a prototype depth sensing camera system for the non-contact monitoring of respiratory rate. The system achieved good performance with high accuracy across a wide range of rates including both clinically important high and low rates.
More Related Videos
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Respiratory Volumes and Capacities I
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...