An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Bond strength of metal-free polyether-ether-ketone knee prostheses compared to metal knee prostheses with bone
Dengxian Wu1,2,3, Qianjin Wang1,2,3, Hung-Kang Tsai1,2,3
1Division of Sports Medicine and Adult Reconstructive Surgery, Department of Orthopedic Surgery, Nanjing Drum Tower Hospital, Affiliated Hospital of Medical School, Nanjing University, Nanjing, Jiangsu, PR China.
Background:
Total knee arthroplasty is the most effective treatment for advanced-stage knee arthritis, and the majority of knee prostheses are made of metal. Nevertheless, metal prostheses still have several problems. The objective of this study is to introduce new metal-free knee prostheses made of polyether-ether-ketone (PEEK) and to compare their cement bond strength with metal prostheses.
Methods:
Twelve sets of knee prostheses were divided into four groups (unloaded PEEK, unloaded Metal, 10 million cycles (MC) PEEK, 10 MC Metal, N = 3 each), and then attached to composite bones using bone cement. Both the 10 MC PEEK and 10 MC Metal groups were subjected to dynamic gait simulations of 10 MC, whereas the other two sets were not. Afterwards, a pull-off strength test was performed on the femoral prostheses and a shear strength test was performed on the tibial prostheses.
Results:
No apparent cracks were observed in the bone cement after subjecting the PEEK and Metal groups to 10 million cycles of dynamic simulation. No statistically significant differences were observed (p > .05) in the strength tests for unloaded PEEK vs. unloaded Metal, 10 MC PEEK vs.10 MC Metal in the femoral pull-off test, and for unloaded PEEK vs. unloaded Metal in the tibial shear test. The shear strength of 10 MC PEEK was significantly lower (p < .05) compared to that of 10 MC Metal.
Conclusions:
By comparing the force analysis of previous investigations on knee prostheses with the failure pattern observed in the PEEK knee prosthesis of this study, which replicates that of the metal prosthesis. We believe that the combination of the peek knee prosthesis with bone cement is reliable. We anticipate that metal-free PEEK knee prostheses will find application in Total Knee Arthroplasty (TKA) in the future, thereby benefiting patients.
More Related Videos
09:08Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Related Concept Videos
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...