Related Experiment Video
Updated: Jul 10, 2025

Computerized Dynamic Posturography for Postural Control Assessment in Patients with Intermittent Claudication
Published on: December 11, 2013
Development of a prediction equation to estimate lower-limb arterial occlusion pressure with a thigh sphygmomanometer
Isaac J Wedig1,2,3, Isaac M Lennox2,3, Erich J Petushek3,4
1School of Health and Human Performance, Northern Michigan University, Marquette, MI, USA.
Introduction:
Previous investigators have developed prediction equations to estimate arterial occlusion pressure (AOP) for blood flow restriction (BFR) exercise. Most equations have not been validated and are designed for use with expensive cuff systems. Thus, their implementation is limited for practitioners.
Purpose:
To develop and validate an equation to predict AOP in the lower limbs when applying an 18 cm wide thigh sphygmomanometer (SPHYG18cm).
Methods:
Healthy adults (n = 143) underwent measures of thigh circumference (TC), skinfold thickness (ST), and estimated muscle cross-sectional area (CSA) along with brachial and femoral systolic (SBP) and diastolic (DBP) blood pressure. Lower-limb AOP was assessed in a seated position at the posterior tibial artery (Doppler ultrasound) using a SPHYG18cm. Hierarchical linear regression models were used to determine predictors of AOP. The best set of predictors was used to construct a prediction equation to estimate AOP. Performance of the equation was evaluated and internally validated using bootstrap resampling.
Results:
Models containing measures of either TC or thigh composition (ST and CSA) paired with brachial blood pressures explained the most variability in AOP (54%) with brachial SBP accounting for majority of explained variability. A prediction equation including TC, brachial SBP, and age showed good predictability (R2 = 0.54, RMSE = 7.18 mmHg) and excellent calibration. Mean difference between observed and predicted values was 0.0 mmHg and 95% Limits of Agreement were ± 18.35 mmHg. Internal validation revealed small differences between apparent and optimism adjusted performance measures, suggesting good generalizability.
Conclusion:
This prediction equation for use with a SPHYG18cm provided a valid way to estimate lower-limb AOP without expensive equipment.
More Related Videos
Related Concept Videos
Assessing Blood pressure in the Leg
Preparation:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: