Related Experiment Video
Updated: Jul 10, 2025

Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
Photon-counting CT urogram: optimal acquisition potential (kV) determination for virtual noncontrast creation
Bari Dane1, Daniel Freedman2, Kun Qian3
1Department of Radiology, NYU Langone Health, 660 1st Avenue, New York, NY, 10016, USA. Bari.Dane@nyulangone.org.
Purpose:
To quantitatively and qualitatively compare the degree of iodine removal in the collecting system from PCCT urographic phase-derived virtual noncontrast (VNC) images obtained at 140 kV versus 120 kV.
Methods:
A retrospective PACS search identified adult patients (>18 years) who underwent a PCCT urogram for hematuria from 4/2022 to 4/2023 with available urographic phase-derived VNC images in PACS. Tube voltage (120 kV, 140 kV), body mass index, CTDIvol, dose length product (DLP), and size-specific dose estimate (SSDE) were recorded. Hounsfield Unit (HU) in both renal pelvises and the urinary bladder on urographic-derived VNC were recorded. Three radiologists qualitatively assessed the degree of iodine removal (renal pelvis, urinary bladder) and diagnostic confidence for urinary stone detection. Continuous variables were compared for 140 kV versus 120 kV with the Wilcoxon rank sum test. A p < .05 indicated statistical significance.
Results:
63 patients (34 male; median (Q1, Q3) age: 30 (26, 34) years; 140 kV/120 kV: 30 patients/33 patients) were included. BMI, CTDIvol, DLP, and SSDE were not different for 140 kV and 120 kV (all p > .05). Median (Q1, Q3) collecting system HU (renal pelvis and bladder) was 0.9 (- 3.6, 4.4) HU at 140 kV and 10.5 (3.6, 26.7) HU at 120 kV (p = .04). Diagnostic confidence for urinary calculi was 4.6 [1.1] at 140 kV and 4.1 [1.4] at 120 kV (p = .005). Diagnostic confidence was 5/5 (all readers) in 82.2% (74/90) at 140 kV and 59.6% (59/99) at 120 kV (p < .001).
Conclusion:
PCCT urographic phase-derived VNC images obtained at 140 kV had better collecting system iodine removal than 120 kV with similar patient radiation exposure. With excellent PCCT urographic phase iodine removal at 140 kV, consideration can be made to utilize a single-phase CT urogram in young patients.