Related Experiment Video
Updated: Jul 10, 2025

Retinal Vascular Reactivity as Assessed by Optical Coherence Tomography Angiography
Published on: March 26, 2020
Influence of different noninvasive oxygenation support devices on tidal volume
Anne-Fleur Haudebourg1,2, Tommaso Maraffi3, Samuel Tuffet4,5
1Service de Médecine Intensive Réanimation, Assistance Publique-Hôpitaux de Paris, CHU Henri Mondor - Albert Chenevier, 51, Avenue du Maréchal de Lattre de Tassigny, 94010, Créteil Cedex, France. annefleur.maignant@aphp.fr.
Background:
Multiple devices are available for noninvasive oxygenation support, including non-rebreather oxygen mask (O2-mask), high-flow oxygen through nasal cannula (HFNC), continuous positive airway pressure (CPAP), mask noninvasive ventilation (Mask-NIV) and helmet NIV (Helmet-NIV). As tidal volume is a key determinant of efficacy and safety during ventilatory support, we assessed whether it was influenced by the type of noninvasive oxygenation device.
Methods:
A bench study using a manikin with a realistic face connected to a lung simulator was performed. Six conditions were assessed: no device, O2-mask, HFNC, CPAP, Mask-NIV and Helmet-NIV. Three respiratory mechanics were simulated (normal, obstructive, restrictive), at three simulated efforts (low, moderate, respiratory distress). Flow was recorded at the lung simulator inlet and mouth pressure into the manikin mouth. The same devices were evaluated on healthy volunteers with tidal volume assessed by electrical impedance tomography (EIT).
Results:
Tidal volume was significantly influenced by oxygenation devices in bench model. As compared to O2-mask, HFNC and CPAP delivered significantly lower tidal volumes (440 ± 352 mL, 414 ± 333 mL and 377 ± 297 mL, respectively), while Mask-NIV or Helmet-NIV were associated with significantly higher tidal volumes (690 ± 321 mL and 652 ± 366 mL, respectively). Tidal volume was strongly correlated with the specific effect of each device on mouth pressure during inspiration: HFNC and CPAP were characterized by a negative PTPmouth (- 0.3 [- 0.8 to - 0.2] and - 0.7 [- 2.2 to - 0.5] cmH2O.sec/cycle, respectively), while Helmet-NIV and Mask-NIV were associated with a positive PTPmouth (4.5 [4.1-4.6] and 6.1 [5.9-7.1] cmH2O.sec/cycle, respectively). Tidal volume was also significantly influenced by oxygenation devices in healthy volunteers, with similar tidal volumes between O2-mask and CPAP (644 [571-764] and 648 [586-770] mL) but higher with HFNC, Mask-NIV and Helmet-NIV (819 [609-918], 1110 [661-1305] and 1086 [833-1243] mL).
Conclusions:
Tidal volume is significantly influenced by noninvasive oxygenation support devices, with a strong correlation with the pressure variation generated into the mouth during inspiration. NIV was associated with the highest tidal volumes and CPAP with the lowest ones. Clinical studies are needed to clarify the clinical implications of these effects.
More Related Videos
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Mechanical Ventilation I: Indication and Settings