Wearable in-ear pulse oximetry validly measures oxygen saturation between 70% and 100%: A prospective agreement study
Catherina Ab Bubb1,2, Michael Weber3, Nadine Kretsch1
1Technical University of Munich, School of Medicine and Health, Department of Anaesthesiology and Intensiv Care Medicine, Munich, Germany.
Objectives:
Postoperative monitoring outside intensive and post-anaesthesia care units is seldom, partly due to lack of suitable and approved systems. We therefore aim to validate the oxygen saturation (SpO2) and pulse rate measurement of the in-ear sensor c-med° alpha with a reference pulse oximeter.
Methods:
This prospective agreement study was conducted in 12 healthy (ASA 1) adult (18-50 years) volunteers according to the EN ISO 80601-2-61. The sitting volunteers were equipped with the finger pulse oximeter Rad-5 and two c-med° alpha sensors in each ear. The inspiratory oxygen content was reduced via a tight-fitting breathing mask to achieve five defined plateaus with stable SpO2 between 99% and 70%. The deviation of the SpO2 and pulse rate measurements of the c-med° alpha from those of the Rad-5 was calculated using the mean square error (Arms). Bias and limits of agreement between both devices were calculated using the Bland-Altman technique. The precision was compared based on the repeatability coefficients.
Results:
The c-med° alpha measured SpO2 had an Arms = 1.9% relative to the Rad-5, a non-significant bias (-0.1% (-0.2% to 0.0%)), levels of agreement from -4.0% to 3.8%, and the same repeatability coefficient (0.8% vs. 0.8%). The c-med° alpha measured pulse rate did not deviate from the one measured with the certified finger pulse oximeter (bias: 0.1 min-1 (0 to 0.1 min-1), level of agreement: -3.6 to 3.7 min-1, Arms: 1.8 min-1).
Conclusions:
The c-med° alpha fulfils the EN ISO 80601-2-61 standard and is sufficiently accurate for measuring SpO2 and pulse rate in healthy adults at rest.
Trial Registration:
EUDAMED No. CIV-21-03-036033.
More Related Videos
10:45A Community-based Stress Management Program: Using Wearable Devices to Assess Whole Body Physiological Responses in Non-laboratory Settings
Published on: January 22, 2018
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...