Measurement Accuracy When Using Spot Vision Screener With or Without Cycloplegia in Young Adults
Shunya Tatara1,2, Fumiatsu Maeda1,3, Hokuto Ubukata1
1Department of Orthoptics and Visual Sciences, Faculty of Medical Technology, Niigata University of Health and Welfare, Niigata, Japan.
Clinical Ophthalmology (Auckland, N.Z.)
|November 29, 2023
Summary
The Spot Vision Screener (SVS) is less accurate for cycloplegic refraction measurements in myopia compared to autorefractometers. Autorefractometers are preferred for more reliable myopia assessment under cycloplegia.
Area of Science:
- Ophthalmology
- Optometry
- Vision Science
Background:
- Accurate refractive error measurement is crucial for myopia management.
- The Spot Vision Screener (SVS) is a handheld photorefractor used for vision screening.
- Uncertainty exists regarding the accuracy of SVS for cycloplegic refraction, particularly in myopic individuals.
Purpose of the Study:
- To evaluate the accuracy of the Spot Vision Screener (SVS) for measuring cycloplegic refraction in healthy subjects.
- To compare SVS measurements with autorefractometers and subjective refraction under cycloplegia.
Main Methods:
- Forty-nine healthy subjects participated in the study.
- Refraction was measured using SVS, a table-mounted autorefractometer (RT7000), and a handheld autorefractometer (Retinomax Screeen).
- Measurements were taken under both noncycloplegic and cycloplegic conditions (1% cyclopentolate), with subjective refraction obtained using a cross-cylinder.
Main Results:
- Under cycloplegia, SVS measurements showed a significantly less myopic spherical equivalent (SE) compared to autorefractometers.
- SVS exhibited both fixed (more hyperopic) and proportional errors when compared to subjective refraction under cycloplegia.
- Autorefractometers provided more myopic SE values than SVS in the cycloplegic condition.
Conclusions:
- Autorefractometers are preferable to photorefractors like SVS for cycloplegic refraction measurements.
- The accuracy of SVS for cycloplegic refraction in myopia requires further consideration.
- Clinical decisions regarding myopia management may benefit from using autorefractometers for cycloplegic refractions.