International Council for Standardization in Haematology Field Study Evaluating Optimal Interpretation Methods for
Robert C Gosselin1, Gary W Moore2,3, Geoffrey W Kershaw4
1From the Hemostasis & Thrombosis Center, University of California, Davis Health System, Sacramento (Gosselin).
Context.—:
The prothrombin time (PT) and activated partial thromboplastin time (APTT) are screening tests used to detect congenital or acquired bleeding disorders. An unexpected PT and/or APTT prolongation is often evaluated using a mixing test with normal plasma. Failure to correct ("noncorrection") prolongation upon mixing is attributed to an inhibitor, whereas "correction" points to factor deficiency(ies).
Objective.—:
To define an optimal method for determining correction or noncorrection of plasma mixing tests through an international, multisite study that used multiple PT and APTT reagents and well-characterized plasma samples.
Design.—:
Each testing site was provided 22 abnormal and 25 normal donor plasma samples, and mixing studies were performed using local PT and APTT reagents. Mixing study results were evaluated using 11 different calculation methods to assess the optimal method based on the expected interpretation for factor deficiencies (correction) and noncorrection (inhibitor effect). Misprediction, which represents the failure of a mixing study interpretation method, was assessed.
Results.—:
Percentage correction was the most suitable calculation method for interpreting PT mixing test results for nearly all reagents evaluated. Incubated PT mixing tests should not be performed. For APTT mixing tests, percentage correction should be performed, and if the result indicates a factor deficiency, this should be confirmed with the subtraction III calculation where the normal pooled plasma result (run concurrently) is subtracted from the mixing test result with correction indicated by a result of 0 or less. In general, other calculation methods evaluated that performed well in the identification of factor deficiency tended to have high misprediction rates for inhibitors and vice versa.
Conclusions.—:
No single method of mixing test result calculation was consistently successful in accurately distinguishing factor deficiencies from inhibitors, with between-reagent and between-site variability also identified.
Related Concept Videos
Blood Transfusion and Agglutination
History
The history of blood transfusion dates back to the 17th century, when early attempts were made in animals. In 1818 James Blundell, a British doctor, performed the first successful human blood transfusion. Later in 1900, Karl...
Fast Reactions
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...