A novel, benchtop model for quantitative analysis of resistance in ventricular catheters
Pranav Gopalakrishnan1, Ahmad Faryami2, Carolyn A Harris3
1Department of Medical Education, Wayne State University School of Medicine, Detroit, MI, United States of America.
Introduction:
The mechanisms of catheter obstruction are still poorly understood, but the literature suggests that resistance to fluid flow plays a significant role. We developed and assessed a gravity-driven device that measures flow through ventricular catheters. We used this device to quantitatively analyze the resistances of unused ventricular catheters used in the treatment of hydrocephalus; failed hydrocephalus catheters from our catheter biorepository were also evaluated quantitatively.
Methods:
Catheters of three manufacturing companies were inserted into the benchtop model, which records time, flow rate, and pressure data using sensors. The relative resistances of catheters across six design models were evaluated. Experiments were performed to evaluate changes in the relative resistance of a catheter when the catheter's holes were progressively closed. The relative resistance of explanted catheters from our catheter biorepository was also measured.
Results:
Experimental results showed significant differences (P<0.05) between the relative resistances of different catheter models just after being removed from their packaging. A non-linear trend of increasing resistance was observed in experiments on catheters with artificially obstructed holes. Data from five individual benchtop models were compared, and the differences in measured data between the models were found to be negligible. A significant increase (P < 0.05) in relative resistance was observed in explanted catheters.
Conclusion:
The current study sought to propose a novel in-vitro model and use it to examine data on differences in relative resistance among catheter models. From these experiments, we can rapidly correlate clinical patient cohorts to identify mechanisms of luminal shunt obstruction.