Shielding block mounting with double-stick tape.

R J Barish

    International Journal of Radiation Oncology, Biology, Physics
    |January 1, 1987
    PubMed
    Summary

    A novel double-stick adhesive tape system offers a simpler and more reliable method for mounting radiotherapy shielding blocks. This technique avoids dose inaccuracies associated with traditional screw-and-nut systems, proving effective in clinical use.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Health physics and aviation: solar cycle 23 (1996-2008).

    Health physics·2009
    Same author

    Radiation protection analysis for a rare mineral specimen.

    Health physics·2001
    Same author

    NOAA space weather alerts. National Oceanic and Atmospheric Administration.

    Health physics·2000
    Same author

    In-flight radiation: counseling patients about risk.

    The Journal of the American Board of Family Practice·1999
    Same author

    On designing room shielding for total-body irradiation.

    Health physics·1996
    Same author

    Health physics and aviation: 1990-1994.

    Health physics·1995

    Area of Science:

    • Medical Physics
    • Radiation Oncology
    • Radiotherapy Equipment

    Background:

    • Traditional methods for mounting radiotherapy shielding blocks, such as screw-and-nut systems, can introduce dose perturbations.
    • These perturbations arise from necessary holes or slots in the blocking tray, potentially affecting treatment accuracy.

    Purpose of the Study:

    • To introduce and evaluate an alternative system for mounting radiotherapy shielding blocks.
    • To assess the feasibility and reliability of using double-stick adhesive tape for this purpose.

    Main Methods:

    • A novel system utilizing double-stick adhesive tape was developed for mounting radiotherapy shielding blocks.
    • The system was implemented and evaluated over a period of more than two years in a clinical setting.

    Main Results:

    • The double-stick adhesive tape system proved to be easy to use.
    • This method effectively eliminated dose perturbations caused by holes or slots in blocking trays.
    • Clinical experience exceeding two years confirmed the system's reliability.

    Conclusions:

    • Double-stick adhesive tape presents a viable, reliable, and simpler alternative for mounting radiotherapy shielding blocks.
    • This method enhances treatment accuracy by preventing dose perturbations associated with conventional mounting systems.