Optimisation of the air fraction correction for lung PET/CT: addressing resolution mismatch
Francesca Leek1,2, Cameron Anderson3, Andrew P Robinson4,5,6
1Institute of Nuclear Medicine, University College London Hospitals NHS Trust, London, UK. francesca.leek.09@ucl.ac.uk.
Background:
Increased pulmonary [Formula: see text]F-FDG metabolism in patients with idiopathic pulmonary fibrosis, and other forms of diffuse parenchymal lung disease, can predict measurements of health and lung physiology. To improve PET quantification, voxel-wise air fractions (AF) determined from CT can be used to correct for variable air content in lung PET/CT. However, resolution mismatches between PET and CT can cause artefacts in the AF-corrected image.
Methods:
Three methodologies for determining the optimal kernel to smooth the CT are compared with noiseless simulations and non-TOF MLEM reconstructions of a patient-realistic digital phantom: (i) the point source insertion-and-subtraction method, [Formula: see text]; (ii) AF-correcting with varyingly smoothed CT to achieve the lowest RMSE with respect to the ground truth (GT) AF-corrected volume of interest (VOI), [Formula: see text]; iii) smoothing the GT image to match the reconstruction within the VOI, [Formula: see text]. The methods were evaluated both using VOI-specific kernels, and a single global kernel optimised for the six VOIs combined. Furthermore, [Formula: see text] was implemented on thorax phantom data measured on two clinical PET/CT scanners with various reconstruction protocols.
Results:
The simulations demonstrated that at [Formula: see text] iterations (200 i), the kernel width was dependent on iteration number and VOI position in the lung. The [Formula: see text] method estimated a lower, more uniform, kernel width in all parts of the lung investigated. However, all three methods resulted in approximately equivalent AF-corrected VOI RMSEs (<10%) at [Formula: see text]200i. The insensitivity of AF-corrected quantification to kernel width suggests that a single global kernel could be used. For all three methodologies, the computed global kernel resulted in an AF-corrected lung RMSE <10% at [Formula: see text]200i, while larger lung RMSEs were observed for the VOI-specific kernels. The global kernel approach was then employed with the [Formula: see text] method on measured data. The optimally smoothed GT emission matched the reconstructed image well, both within the VOI and the lung background. VOI RMSE was <10%, pre-AFC, for all reconstructions investigated.
Conclusions:
Simulations for non-TOF PET indicated that around 200i were needed to approach image resolution stability in the lung. In addition, at this iteration number, a single global kernel, determined from several VOIs, for AFC, performed well over the whole lung. The [Formula: see text] method has the potential to be used to determine the kernel for AFC from scans of phantoms on clinical scanners.
More Related Videos
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...