Validation of a Portable Wireless Force Platform System To Measure Ground Reaction Forces During Various Tasks
Vasileios Mylonas1, Christos Chalitsios1, Thomas Nikodelis1
1Physical Education and Sport Science Aristotle University of Thessaloniki.
Background:
Force platforms are widely used in biomechanics to measure ground reaction forces (GRF) during various human movements. However, traditional force plates are not easily used outside a research lab. To overcome this issue, researchers and manufacturers are developing low-cost portable force platforms that can be used in a variety of settings, including outdoors.
Purpose:
To validate the kinetic data obtained from a pair of portable K-Deltas force platforms compared to gold standard platforms fixed in the lab and to examine the measurement reliability between this pair of portable force platforms.
Methods:
Force-time curves from known masses, countermovement vertical jumps, and balance tests were used to assess validity of K-Deltas using a pair of Bertec force plates as a gold standard and between the K-Deltas pair of plates. Bland-Altman plots were used to evaluate the differences between K-Deltas and Bertec force plates. For the assessment of countermovement vertical jumps, impulse, peak rate of force development and peak force were calculated for both instruments and checked for agreement between instruments. Three young adults (2 male, 1 female, 25.4±0.83 years) participated in the study.
Results:
The percentage of Bland-Altman plot point within the limits of agreement was 94.59 % for the comparison between K-Deltas and Bertec and 94.83% between the pair of K-Deltas.
Conclusion:
The results show that the portable force platforms could be utilized successfully for assessing pertinent parameters in clinical and sports biomechanics. The findings suggest that portable force platforms can be used as an alternative to traditional laboratory equipment for field assessment, providing significant improvements compared to the past.
Level Of Evidence:
Level 3.
More Related Videos
06:21Postural Organization of Gait Initiation for Biomechanical Analysis Using Force Platform Recordings
Published on: July 26, 2022
08:58Use of a Foot-Induced Digitally Controlled Resistance Device for Functional Magnetic Resonance Imaging Evaluation in Patients with Foot Paresis
Published on: July 7, 2023
Related Concept Videos
Static Equilibrium - II
Apparent Weight
Consider a person standing on a bathroom scale inside an elevator. If the scale is accurate at rest, its reading equals the...
Measuring Acceleration Due to Gravity
A simple pendulum can be described as a point mass and a string. Meanwhile, a physical pendulum is any object whose oscillations are similar to a simple pendulum, but cannot be modeled as a point mass on a string because its mass is distributed over a larger area. The behavior of a physical pendulum can be modeled using the principles of...
Vebe Test
To conduct the test, concrete is placed into the slump cone. The concrete is filled in layers and...
Non-destructive Tests for Concrete Strength