Related Experiment Video
Updated: Jun 30, 2026

Nasal Brushing Sampling and Processing Using Digital High Speed Ciliary Videomicroscopy – Adaptation for the COVID-19 Pandemic
Published on: November 7, 2020
Device comparison study to measure nasal nitric oxide in relation to primary ciliary dyskinesia
Nils Oskar Jõgi1, Karin Ersson1, Kjell Alving2
1Department of Medical Sciences, Uppsala University, Uppsala, Sweden.
Abstract:
Primary ciliary dyskinesia (PCD) is a genetic respiratory disease characterized by chronic cough, recurrent respiratory infections, and rhinosinusitis. The measurement of nasal nitric oxide (nNO) against resistance has been suggested as a sensitive screening method. However, current recommendations argue for the use of expensive, chemiluminescence devices to measure nNO. This study aimed to compare nNO measurement using three different devices in distinguishing PCD patients from healthy controls and cystic fibrosis (CF) patients and to evaluate their diagnostic precision. The study included 16 controls, 16 PCD patients, and 12 CF patients matched for age and sex. nNO measurements were performed using a chemiluminescence device (Eco Medics CLD 88sp), and two devices based on electrochemical sensors (Medisoft FeNO+ and NIOX Vero) following standardized guidelines. Correlation estimation, Bland-Altman, ROC curve, and one-way ANOVA were used to assess device differences and diagnostic performance. Significantly lower nNO output values were observed in PCD and CF patients compared to controls during exhalation against resistance. The correlation analysis showed high agreement among the three devices. ROC curve analysis demonstrated 100% sensitivity and specificity at different cut-off values for all devices in distinguishing PCD patients from controls (optimal cut-offs: EcoMedics 73, Medisoft 92 and NIOX 87 (nl min-1)). Higher nNO output values were obtained with the Medisoft and NIOX devices as compared to the EcoMedics device, with a bias of-19 nl min-1(95% CI: -73-35) and -21 nl min-1(-73-31) accordingly. These findings indicate that all three tested devices can potentially serve as diagnostic tools for PCD if device specific cut-off values are used. This last-mentioned aspect warrants further studies and consideration in defining optimal cut-offs for individual device.
More Related Videos
11:13Collection, Expansion, and Differentiation of Primary Human Nasal Epithelial Cell Models for Quantification of Cilia Beat Frequency
Published on: November 10, 2021
05:32High-speed Video Microscopy Analysis for First-line Diagnosis of Primary Ciliary Dyskinesia
Published on: January 19, 2022