Related Experiment Video
Updated: Jun 21, 2026

A New Murine Model of Endovascular Aortic Aneurysm Repair
Published on: July 7, 2013
Intra-stent aortic wall aneurysm formation after Be-graft covered stent implant
Lisa Bianco1, Micol Rebonato1, Gianfranco Butera1
1Department of Cardiology, Cardiac Surgery and Heart Lung Transplantation, ERN GUARD HEART: Bambino Gesù Hospital and Research Institute, IRCCS, Rome, Italy.
Abstract:
Aortic wall injuries may occur after interventional treatment of aortic coarctation (CoA), especially after balloon angioplasty. We reported on a patient who presented with an intra-stent aneurysm formation after direct stenting of a native near atretic aortic CoA by using a BeGraft Aortic stent. This evidence supports the need to maintain a strict follow-up protocol. A computed tomography scan is mandatory, after covered stent implantation as well, especially in high-risk cases and even in the absence of any immediate apparent complication.