Assessment of Heart Rate Monitoring During Exercise With Smart Wristbands and a Heart Rhythm Patch: Validation and
Tse-Lun Wang1,2, Hao-Yi Wu3, Wei-Yun Wang4
1Division of Trauma and Surgical Critical Care, Department of Surgery, Kaohsiung Medical University Hospital, Kaohsiung Medical University, Kaohsiung City, Taiwan.
Background:
The integration of wearable devices into fitness routines, particularly in military settings, necessitates a rigorous assessment of their accuracy. This study evaluates the precision of heart rate measurements by locally manufactured wristbands, increasingly used in military academies, to inform future device selection for military training activities.
Objective:
This research aims to assess the reliability of heart rate monitoring in chest straps versus wearable wristbands.
Methods:
Data on heart rate and acceleration were collected using the Q-Band Q-69 smart wristband (Mobile Action Technology Inc) and compared against the Zephyr Bioharness standard measuring device. The Lin concordance correlation coefficient, Pearson product moment correlation coefficient, and intraclass correlation coefficient were used for reliability analysis.
Results:
Participants from a Northern Taiwanese medical school were enrolled (January 1-June 31, 2021). The Q-Band Q-69 demonstrated that the mean absolute percentage error (MAPE) of women was observed to be 13.35 (SD 13.47). Comparatively, men exhibited a lower MAPE of 8.54 (SD 10.49). The walking state MAPE was 7.79 for women and 10.65 for men. The wristband's accuracy generally remained below 10% MAPE in other activities. Pearson product moment correlation coefficient analysis indicated gender-based performance differences, with overall coefficients of 0.625 for women and 0.808 for men, varying across walking, running, and cooldown phases.
Conclusions:
This study highlights significant gender and activity-dependent variations in the accuracy of the MobileAction Q-Band Q-69 smart wristband. Reduced accuracy was notably observed during running. Occasional extreme errors point to the necessity of caution in relying on such devices for exercise monitoring. The findings emphasize the limitations and potential inaccuracies of wearable technology, especially in high-intensity physical activities.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes