Related Experiment Video
Updated: Jul 8, 2025

Author Spotlight: Advancing Facial Rejuvenation Therapy with Post-Laser Salicylic Acid Application
Published on: September 27, 2024
Evaluation of a Novel Microfocused Ultrasound with Three-Dimensional Digital Imaging for Facial Tightening: A
Jiafang Zhu1, Yue Han1, Ying Liu2
1Department of Laser and Aesthetic Medicine, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, 200011, China.
Background:
The excellent efficacy is mitigated by the limited safety profile of microfocused ultrasound procedures.
Objective:
We sought to assess the safety and tightening efficacy of a novel microfocused ultrasound.
Methods:
The randomized middle and lower face and submental region of the participants were treated with the novel device using the following transducers: M4.5, D4.5, M3.0, and D3.0. Improvement in paired comparison of pretreatment and posttreatment photographs, three-dimensional (3D) volumetric assessments, skin thickness measured by B-ultrasonography, and skin photoaging parameters were evaluated. Adverse events and patient satisfaction were also recorded.
Results:
A total of 20 participants (20 female) were enrolled. Fourteen of 20 participants (70%) were judged to show clinically significant facial tightening during 3-month follow-up (P < 0.05). The mean volumetric change in the lower face, as quantitatively assessed after 3 months was -0.29 mL compared with +0.42 mL on the control side (P < 0.05). The VAS pain score was 3.00 ± 1.19 without any oral or intramuscular anesthesia.
Conclusions:
A small sample size, lack of clinical scales, and impersonalized treatment parameters. The novel microfocused ultrasound appears to be a safe and effective modality for lower-face tightening.
Clinical Trial Registration Number:
ChiCTR 2200064666.
More Related Videos
08:08Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
05:56Evaluation of the Feasibility, Safety, and Accuracy of an Intraoperative High-intensity Focused Ultrasound Device for Treating Liver Metastases
Published on: January 9, 2019