Robustness of Quantitative Diffusion Metrics from Four Models: A Prospective Study on the Influence of Scan-Rescans,
Jingyu Zhong1, Xianwei Liu1, Yangfan Hu1
1Department of Imaging, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
Background:
Quantitative diffusion metrics provide additional microstructural information of diseases. The robustness of quantitative diffusion metrics should be established before clinical application.
Purpose:
To evaluate the variability and reproducibility of quantitative diffusion MRI metrics.
Study Type:
Prospective.
Population:
14 volunteers (7 men; median age, range, 28, 26-59 years).
Field Strength/Sequence:
3.0-T/Diffusion spectrum imaging.
Assessment:
Brain MRI studies were performed four times per subject: involving different combinations of coil types and voxel sizes. Regions of interest of 13 brain anatomical sites were drawn by one observer twice and another observer once to allow interobserver and intraobserver reproducibility assessment. Twenty-five quantitative metrics were calculated using four diffusion models.
Statistical Tests:
The variability was evaluated with coefficients of variation (CV), and quartile coefficient of dispersion (QCD). The reproducibility was assessed with intraclass correlation coefficient (ICC), and concordance correlation coefficient (CCC). Wilcoxon signed rank test was used to compare the influence of factors on robustness of quantitative diffusion metrics. A two-tailed P < 0.05 was considered statistically significant.
Results:
The variability of quantitative diffusion metrics showed CV of 2.4%-68.2%, and QCD of 0.6%-48.2%, respectively. The reproducibility of scans using 20-channel coils with voxels of 2 × 2 × 2 mm3 and 3 × 3 × 3 mm3, respectively (ICC 0.03-0.84, CCC 0.03-0.84) was significantly worse than that of repeated scans using a 20-channel coil with a voxel size of 2 × 2 × 2 mm3 (ICC of 0.74-0.97, CCC 0.74-0.97) and that of scans using 20- and 64-channel coils, respectively, with a voxel size of 2 × 2 × 2 mm3 (ICC 0.59-0.95, CCC 0.59-0.95). The intraobserver reproducibility (ICC 0.49-0.94, CCC 0.49-0.94) was significantly better than the interobserver reproducibility (ICC 0.28-0.91, CCC 0.28-0.91).
Data Conclusion:
Our study indicated that the voxel size has a greater influence on the reproducibility of quantitative diffusion metrics than scan-rescans and coils. The reproducibility within one observer was higher than that between two observers.
Evidence Level:
2 TECHNICAL EFFICACY: Stage 1.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Imaging Studies IV: Magnetic Resonance Imaging