Elemental analysis of hourly collected air filters with X-ray fluorescence under grazing incidence
Hikari Takahara1, Atsushi Morikawa2, Saori Kitayama3
1X-ray Product Div., Rigaku Corporation, 14-8 Akaoji-cho, Takatsuki, Osaka, 569-1146, Japan. hikari@rigaku.co.jp.
Abstract:
The X-ray fluorescence under grazing incidence condition (XRF-UGI) was applied for the direct analysis of aerosol filters. Particulate matter less than 2.5 microns (PM2.5) was collected hourly on polytetrafluoroethylene filters using a continuous PM monitor with a virtual impactor method. Although the sampling mass is in trace amounts of 5-30 μg, the metallic contents, such as V, Cr, Mn, Fe, Zn, and Pb, can be measured at sub-ng m-3 detection limits. The effects of the non-uniformity and poor flatness of the PM filters were discussed with regard to the measurement repeatability. The relationship between the XRF-UGI intensities and the mass concentrations obtained via conventional X-ray fluorescence (XRF) analysis was confirmed using the fundamental parameter method. Finally, quantification was successfully demonstrated using the XRF-UGI results with the relative sensitivity factors.