Related Experiment Video
Updated: Jun 28, 2026

Biochemical Measurement of Neonatal Hypoxia
Published on: August 24, 2011
Accuracy of transcutaneous bilirubinometry in term infants after phototherapy: a prospective observational study
Charissa Joy1, Taya A Collyer2,3, Kathryn McMahon4
1Department of Paediatrics, Peninsula Health, Frankston, Australia.
Objective:
To test the accuracy of transcutaneous bilirubinometry (TcB) in neonates 12 h after discontinuing phototherapy.
Study Design:
In a prospective study of 91 neonates at ≥35 weeks of gestation, paired measurements of total serum bilirubin (TSB) and TcB were obtained 12 h after discontinuation of phototherapy. TcB measurements were obtained on the uncovered skin of the sternum and the covered skin of the lower abdomen. Bland-Altman plots were used to evaluate agreement between TSB and TcB.
Results:
TcB was found to systematically underestimate TSB on both covered and uncovered skin. The smallest but statistically significant difference between TSB and TcB was found on the covered lower abdomen (-1.03, p < .0001) compared with the uncovered skin of the sternum (-1.44, p < .0001). The correlation between TSB and TcB was excellent on both covered (r = 0.86, p < .001) and uncovered skin (r = 0.90, p < .001). Bland and Altman plots showed poor agreement between TcB and TSB.
Conclusions:
This study demonstrated excellent correlation between TcB and TSB 12 h after phototherapy but poor TcB-TSB agreement. TcB cannot be reliably used in neonates exposed to phototherapy.
Related Concept Videos
Pharmacokinetics in Pediatric Patients: Drug Metabolism
Jaundice