A novel pneumatic drill for bone biopsy under MRI imaging
Elizabeth Fischer1, Mikias Gebremeskel2, Dan Stoianovici3
1Sheikh Zayed Institute for Pediatric Surgical Innovation, Children's National Hospital, Washington, DC, USA. emfischer2@childrensnational.org.
A new MRI-safe pneumatic drill for bone biopsies yielded high-quality tissue samples comparable to standard battery-powered drills. This innovation aims to improve lesion visualization and eliminate radiation exposure during bone biopsy procedures.
Area of Science:
- Orthopedic Surgery
- Medical Device Engineering
- Radiology
Background:
- Current computed tomography (CT)-guided bone biopsies use battery-powered drills, which offer limited soft tissue discrimination and involve ionizing radiation.
- These limitations necessitate the development of advanced imaging and biopsy techniques for improved diagnostic accuracy and patient safety.
Purpose of the Study:
- To develop and evaluate an MRI-safe pneumatic drill as an alternative to traditional CT-guided bone biopsy methods.
- To compare the drilling time and sample quality obtained using the novel pneumatic drill versus a standard battery-powered drill in an animal model.
Main Methods:
- Five porcine rib bones were utilized for the study.
- Each bone specimen was subjected to drilling with both an 11-gauge battery-powered drill and a newly developed pneumatically actuated drill.
- Drilling time for each sample was recorded using a stopwatch.
Main Results:
- All 20 bone samples obtained from both drills exhibited high-quality inner cores and intact cortices.
- The average drilling time for the pneumatic drill was 8.5 ± 2.5 seconds, compared to 7.1 ± 1.4 seconds for the battery-powered drill.
- The pneumatic drill demonstrated slightly longer drilling times, with an average difference of 1.39 seconds.
Conclusions:
- Both the pneumatic and battery-powered drills successfully acquired comparable bone biopsy specimens.
- The pneumatic drill's slightly increased drilling time is considered clinically insignificant.
- The developed MRI-safe pneumatic drill holds promise for enabling radiation-free bone biopsies with enhanced lesion visualization under MRI guidance.