Improved test-retest reliability of R2* and susceptibility quantification using multi-shot multi-echo 3D EPI
Yujia Huang1, Lin Chen2,3, Xu Li2,3
1Advanced Imaging Research Center, UT Southwestern Medical Center, Dallas, Texas, USA.
Purpose:
This study aimed to evaluate the potential of 3D EPI for improving the reliability of -weighted data and quantification of decay rate and susceptibility (χ) compared with conventional gradient-echo (GRE)-based acquisition.
Methods:
Eight healthy subjects in a wide age range were recruited. Each subject received repeated scans for both GRE and EPI acquisitions with an isotropic 1 mm resolution at 3 T. Maps of and χ were quantified, and their interscan differences were used to evaluate the test-retest reliability. Interprotocol differences of and χ between GRE and EPI were also measured voxel by voxel and in selected regions of interest to test the consistency between the two acquisition methods.
Results:
The quantifications of and χ using EPI protocols showed increased test-retest reliability with higher EPI factors up to 5 as performed in the experiment and were consistent with those based on GRE.
Conclusion:
The result suggests that multishot multi-echo 3D EPI can be a useful alternative acquisition method for -weighted MRI and quantification of and χ with reduced scan time, improved test-retest reliability, and similar accuracy compared with commonly used 3D GRE.