Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments
Published on: January 20, 2022
MOCCal: A Multiomic CCS Calibrator for Traveling Wave Ion Mobility Mass Spectrometry
Hannah M Hynds1, Kelly M Hines1
1Department of Chemistry, University of Georgia, 302 East Campus Road, Athens, Georgia 30602, United States.
Abstract:
Ion mobility mass spectrometry (IM-MS) is a rapid, gas-phase separation technology that can resolve ions on the basis of their size-to-charge and mass-to-charge ratios. Since each class of biomolecule has a unique relationship between size and mass, IM-MS spectra of complex biological samples are organized into trendlines that each contain one type of biomolecule (i.e., lipid, peptide, metabolite). These trendlines can aid in the identification of unknown ions by providing a general classification, while more specific identifications require the conversion of IM arrival times to collision cross section (CCS) values to minimize instrument-to-instrument variability. However, the process of converting IM arrival times to CCS values varies between the different IM devices. Arrival times from traveling wave ion mobility (TWIM) devices must undergo a calibration process to obtain CCS values, which can impart biases if the calibrants are not structurally similar to the analytes. For multiomic mixtures, several different types of calibrants must be used to obtain the most accurate CCS values from TWIM platforms. Here we describe the development of a multiomic CCS calibration tool, MOCCal, to automate the assignment of unknown features to the power law calibration that provides the most accurate CCS value. MOCCal calibrates every experimental arrival time with up to three class-specific calibration curves and uses the difference (in Å2) between the calibrated TWCCSN2 value and DTCCSN2 vs m/z regression lines to determine the best calibration curve. Using real and simulated multiomic samples, we demonstrate that MOCCal provides accurately calibrated TWCCSN2 values for small molecules, lipids, and peptides.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
16:40T-wave Ion Mobility-mass Spectrometry: Basic Experimental Procedures for Protein Complex Analysis
Published on: July 31, 2010
Related Concept Videos
Mass Analyzers: Common Types
Mass Spectrometers
MALDI-TOF Mass Spectrometry
Matrix-assisted laser desorption ionization (MALDI) is a commonly...
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
Mass Analyzers: Overview
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview