Serum and cerebrospinal fluid neurofilament light chains measured by SIMOA™, Ella™, and Lumipulse™ in multiple
D Vecchio1, C Puricelli2, S Malucchi3
1Neurology Unit, Department of Translational Medicine, Maggiore della Carità University Hospital, University of Piemonte Orientale, Corso Mazzini 18, Novara 28100, Italy.
Background:
Neurofilament light chains (NfL) are cytoskeletal biomarkers of axonal damage, about 40-fold higher in cerebrospinal fluid (CSF) compared to serum, and requiring ultrasensitive techniques to be measured in this latter fluid.
Objectives:
To compare CSF and serum NfL levels in multiple sclerosis (MS) patients using different platforms.
Methods:
60 newly diagnosed relapsing-remitting MS patients (38 females; median age: 36.5 years, range: 15-60) were enrolled before steroid or disease-modifying treatments. CSF and serum NfL were measured with: the commercial Ella™ microfluidic platform (Bio-Techne), the Lumipulse™ Chemiluminescent Enzyme ImmunoAssay (Fujirebio), and the SIMOA™ on the SR-X instrument using NF-light assays (Quanterix).
Results:
CSF and serum NfL absolute levels strongly correlated between assays, although being more elevated with Ella™. Passing-Bablok regression showed high agreement in measuring CSF NfL between assays (with greater proportional difference using Ella™), and very high agreement for serum comparing SIMOA™ and Lumipulse™. Similarly, the Bland-Altman comparison evidenced lower biases for Lumipulse™ for both fluids.
Conclusions:
CSF and serum NfL in naïve MS patients are reliably measured with all assays. Although not interchangeable, SIMOA™ and Lumipulse™ showed high agreement for serum and CSF values.
More Related Videos
08:05Bioluminescence and Near-infrared Imaging of Optic Neuritis and Brain Inflammation in the EAE Model of Multiple Sclerosis in Mice
Published on: March 1, 2017
08:17Author Spotlight: Unveiling the Pathway Linking Obesity to Autoimmune Inflammation in Multiple Sclerosis
Published on: February 23, 2024
Related Concept Videos
Myasthenia Gravis: Diagnostic Tests
The edrophonium test is a diagnostic tool for myasthenia gravis. It involves...
Multiple Sclerosis l: Introduction