Inter and intradevice assessment of microperimetry testing in aging eyes

Leonard M Coulibaly1, Hamza Mohamed2, Philipp Fuchs1

  • 1Department of Ophthalmology and Optometry, Vienna Clinical Trial Centre (VTC), Medical University of Vienna, Währinger Gürtel 18-20, 1090, Vienna, Austria.

Scientific Reports
|January 10, 2024
PubMed

Insights

Pointwise retinal sensitivity (PWS) repeatability was assessed in healthy aging adults using two microperimetry (MP) devices. Older age and the central macula showed reduced PWS repeatability, impacting clinical study outcomes.

Area of Science:

  • Ophthalmology
  • Medical Imaging
  • Gerontology

Background:

  • Microperimetry (MP) combines retinal imaging and sensitivity testing, gaining importance as a clinical study outcome measure.
  • Assessing the repeatability of MP is crucial for its reliable application in research and clinical settings.

Purpose of the Study:

  • To investigate the repeatability of pointwise retinal sensitivity (PWS) measurements using two advanced microperimetry devices (MP-3 and MAIA).
  • To evaluate the influence of age and macular region on PWS repeatability in a healthy aging population.

Main Methods:

  • Two successive MP examinations were conducted on 20 healthy aging subjects (mean age 70 years) using MP-3 and MAIA devices in a randomized order.
  • Pointwise retinal sensitivity (PWS) repeatability was analyzed using Bland-Altmann coefficients of repeatability (CoR) across different macular regions and age groups.

Main Results:

  • Mean CoR values were ±4.61 dB for MAIA and ±4.55 dB for MP-3.
  • Lower PWS repeatability was observed in the central millimeter on MAIA (p=0.005) and with increasing age for both devices (p=0.003).
  • Good intra-device correlation was found (MAIA: 0.79; MP-3: 0.72), with moderate inter-device correlation (0.6).

Conclusions:

  • Older age and the foveal region are associated with worse pointwise retinal sensitivity repeatability in microperimetry.
  • These findings highlight factors influencing the reliability of MP measurements in aging populations and for clinical studies.