Related Experiment Video
Updated: Jul 6, 2025

23:56
Comprehensive & Cost Effective Laboratory Monitoring of HIV/AIDS: an African Role Model
Published on: October 31, 2010
16.7K
Performance Evaluation of Microscanner Plus, an Automated Image-Based Cell Counter, for Counting CD4+ T Lymphocytes
Woong Sik Jang1, Junmin Lee2, Seoyeon Park3
1Departments of Emergency Medicine, College of Medicine, Korea University Guro Hospital, Seoul 08308, Republic of Korea.
Diagnostics (Basel, Switzerland)
|January 11, 2024
Summary
A new automated cell counter, the Microscanner Plus, accurately measures CD4+ T lymphocyte counts. This device offers a simpler, cost-effective alternative to flow cytometry for monitoring HIV patients, especially in resource-limited settings.
Area of Science:
- Immunology
- Medical Technology
- Virology
Background:
- Flow cytometry is the standard for CD4+ T lymphocyte counts in HIV monitoring.
- Resource-limited settings face challenges accessing flow cytometry technology.
- There is a need for simpler, cost-effective CD4+ T cell counting methods.
Purpose of the Study:
- To evaluate the Microscanner Plus, an automated image-based cell counter, for CD4+ T lymphocyte quantification.
- To compare the performance of the Microscanner Plus against standard flow cytometry.
- To assess the suitability of the Microscanner Plus for HIV patient monitoring.
Main Methods:
- The Microscanner Plus was used to count CD4+ T lymphocytes in samples from 47 HIV-infected patients and 87 healthy individuals.
- Results were compared with those obtained from a flow cytometer.
- Precision and reproducibility were assessed using CV% (coefficient of variation).
Main Results:
- The Microscanner Plus demonstrated high precision and reproducibility, with all CV% below 10%.
- The lowest detectable CD4 count was 15.27 cells/µL.
- A very high correlation coefficient (R=0.9906, p<0.0001) was observed between the Microscanner Plus and flow cytometry for CD4 counts in 134 clinical samples.
Conclusions:
- The automated Microscanner Plus exhibits acceptable analytical performance for CD4+ T lymphocyte counting.
- This device presents a viable and potentially valuable tool for monitoring HIV patients in resource-limited environments.
- The Microscanner Plus offers a promising alternative to traditional flow cytometry for CD4+ T cell enumeration.