Improvement in Multi-Angle Plane Wave Image Quality Using Minimum Variance Beamforming with Adaptive Signal Coherence
Che-Chou Shen1, Chun-Lin Huang1
1Department of Electrical Engineering, National Taiwan University of Science and Technology, Taipei 106335, Taiwan.
Sensors (Basel, Switzerland)
|January 11, 2024
Summary
The adaptive delay-multiply-and-sum (ADMAS) method improves ultrasound image quality by adjusting signal coherence locally. This new technique enhances resolution and contrast without sacrificing background speckle quality, outperforming conventional methods.
Area of Science:
- Medical Imaging
- Ultrasound Technology
- Signal Processing
Background:
- Conventional coherent plane wave compounding (CPWC) in ultrasound imaging offers limited quality due to standard delay-and-sum beamforming.
- Delay-multiply-and-sum (DMAS) enhances image quality by incorporating signal coherence but often degrades background speckle.
- Existing DMAS methods apply a global coherence value, impacting overall image quality and contrast-to-noise ratio (CNR).
Purpose of the Study:
- To introduce an adaptive DMAS (ADMAS) algorithm for ultrasound multi-angle plane wave imaging.
- To maintain background speckle quality and CNR by adapting signal coherence to local image characteristics.
- To combine ADMAS with minimum variance (MV) beamforming for improved image resolution.
Main Methods:
- Developed the adaptive DMAS (ADMAS) algorithm, making signal coherence dependent on local image regions.
- Integrated ADMAS with minimum variance (MV) beamforming, optimizing MV estimation along the plane wave transmit angle (Tx).
- Evaluated the TxMV-ADMAS beamforming method using the PICMUS dataset for multi-angle plane wave imaging.
Main Results:
- TxMV-ADMAS significantly improved image quality compared to CPWC.
- While conventional DMAS improved resolution (0.57mm to 0.24mm) and contrast (27.0dB to 38.0dB), it reduced CNR (12.8 to 11.3).
- The proposed ADMAS algorithm restored CNR to CPWC levels while maintaining enhanced resolution and contrast.
Conclusions:
- The proposed TxMV-ADMAS beamforming method effectively enhances ultrasound image resolution and contrast.
- ADMAS preserves background speckle quality and CNR, overcoming limitations of conventional global coherence methods.
- This adaptive approach offers superior image quality for multi-angle plane wave ultrasound imaging.
Keywords:
adaptive delay-multiply-and-sumdelay-multiply-and-summinimum varianceplane wave imagingsignal coherenceMore Related Videos
Related Concept Videos
Linear Approximation in Frequency Domain
91
Linear systems are characterized by two main properties: superposition and homogeneity. Superposition allows the response to multiple inputs to be the sum of the responses to each individual input. Homogeneity ensures that scaling an input by a scalar results in the response being scaled by the same scalar.
In contrast, nonlinear systems do not inherently possess these properties. However, for small deviations around an operating point, a nonlinear system can often be approximated as linear....
In contrast, nonlinear systems do not inherently possess these properties. However, for small deviations around an operating point, a nonlinear system can often be approximated as linear....
91
Linear Approximation in Time Domain
83
Nonlinear systems often require sophisticated approaches for accurate modeling and analysis, with state-space representation being particularly effective. This method is especially useful for systems where variables and parameters vary with time or operating conditions, such as in a simple pendulum or a translational mechanical system with nonlinear springs.
For a simple pendulum with a mass evenly distributed along its length and the center of mass located at half the pendulum's length,...
For a simple pendulum with a mass evenly distributed along its length and the center of mass located at half the pendulum's length,...
83