Optimization of Ultrasound Phase-Encoded Multiplane Wave Tissue Harmonic Imaging
Che-Chou Shen1, Ching-Che Chiu1, You-Lin Chu1
1National Taiwan University of Science and Technology, Taipei, Taiwan.
Ultrasonic Imaging
|March 31, 2026
Summary
Hadamard coding with a rectangular waveform offers optimal phase-encoded multiplane wave tissue harmonic imaging. This configuration balances improved signal-to-noise ratio and reduced artifacts for enhanced diagnostic accuracy.
Area of Science:
- Medical imaging
- Ultrasound technology
- Signal processing
Background:
- Tissue harmonic imaging (THI) offers better resolution than conventional imaging but has low signal-to-noise ratio (SNR).
- Phase-encoded excitation can improve THI's SNR, but signal distortion can cause artifacts.
- Optimizing coding schemes and waveforms is crucial for accurate THI.
Purpose of the Study:
- To compare different coding matrices (Hadamard, S-sequence, Golay) and bit waveforms for phase-encoded multiplane wave (MW) tissue harmonic imaging.
- To identify the optimal configuration for maximizing SNR and minimizing artifacts in THI.
- To validate simulation and experimental findings on coding schemes and waveforms.
Main Methods:
- Simulations and experiments were used to analyze coding schemes and bit waveforms.
- Phase-encoded tissue harmonic imaging with multiplane wave transmission was employed.
- Hadamard, S-sequence, and orthogonal Golay coding matrices were evaluated with different bit waveforms.
Main Results:
- Bipolar codes (Hadamard, Golay) were susceptible to phase distortion, causing axial artifacts.
- Hadamard coding avoided artifacts by adjusting plane wave angles, while S-sequence showed no distortion but lower SNR gain.
- Rectangular waveforms improved SNR but caused spectral distortion, especially with Golay, leading to artifacts and reduced contrast.
Conclusions:
- Hadamard coding with a rectangular waveform and selective compounding is optimal for phase-encoded MW tissue harmonic imaging.
- This configuration provides higher SNR than S-sequence and better image quality than orthogonal Golay.
- The findings offer guidance for improving THI performance and diagnostic capabilities.