Report from the HarmoSter study: different LC-MS/MS androstenedione, DHEAS and testosterone methods compare well;
Flaminia Fanelli1, Mirko Peitzsch2, Stephen Bruce3
1Department of Medical and Surgical Sciences, Endocrinology Research Group, Center for Applied Biomedical Research, Alma Mater Studiorum University of Bologna, Bologna, Italy.
Objectives:
Current liquid chromatography-tandem mass spectrometry (LC-MS/MS) applications for circulating androgen measurements are technically diverse. Previously, variable results have been reported for testosterone. Data are scarce for androstenedione and absent for dehydroepiandrosterone sulfate (DHEAS). We assessed the agreement of androstenedione, DHEAS and testosterone LC-MS/MS measurements among nine European centers and explored benefits of calibration system unification.
Methods:
Androgens were measured twice by laboratory-specific procedures in 78 patient samples and in EQA materials. Results were obtained by in-house and external calibration. Intra- and inter-laboratory performances were valued.
Results:
Intra-laboratory CVs ranged between 4.2-13.2 % for androstenedione, 1.6-10.8 % for DHEAS, and 4.3-8.7 % and 2.6-7.1 % for female and male testosterone, respectively. Bias and trueness in EQA materials were within ±20 %. Median inter-laboratory CV with in-house vs. external calibration were 12.0 vs. 9.6 % for androstenedione (p<0.001), 7.2 vs. 4.9 % for DHEAS (p<0.001), 6.4 vs. 7.6 % for female testosterone (p<0.001) and 6.8 and 7.4 % for male testosterone (p=0.111). Median bias vs. all laboratory median with in-house and external calibration were -13.3 to 20.5 % and -4.9 to 18.7 % for androstenedione, -10.9 to 4.8 % and -3.4 to 3.5 % for DHEAS, -2.7 to 6.5 % and -11.3 to 6.6 % for testosterone in females, and -7.0 to 8.5 % and -7.5 to 11.8 % for testosterone in males, respectively.
Conclusions:
Methods showed high intra-laboratory precision but variable bias and trueness. Inter-laboratory agreement was remarkably good. Calibration system unification improved agreement in androstenedione and DHEAS, but not in testosterone measurements. Multiple components, such as commutability of calibrators and EQA materials and internal standard choices, likely contribute to inter-laboratory variability.
More Related Videos
08:40Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments
Published on: January 20, 2022
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
High-Resolution Mass Spectrometry (HRMS)
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...