Related Experiment Video
Updated: Jul 5, 2025

Studying the Protein Quality Control System of D. discoideum Using Temperature-controlled Live Cell Imaging
Published on: December 2, 2016
A tool for live-cell confocal imaging of temperature-dependent organelle dynamics
Keiko Midorikawa1, Yutaka Kodama1
1Center for Bioscience Research and Education, Utsunomiya University, 350 Mine, Utsunomiya, Tochigi 321-8505, Japan.
Abstract:
Intracellular organelles alter their morphology in response to ambient conditions such as temperature to optimize physiological activities in cells. Observing organelle dynamics at various temperatures deepens our understanding of cellular responses to the environment. Confocal laser microscopy is a powerful tool for live-cell imaging of fluorescently labeled organelles. However, the large contact area between the specimen and the ambient air on the microscope stage makes it difficult to maintain accurate cellular temperatures. Here, we present a method for precisely controlling cellular temperatures using a custom-made adaptor that can be installed on a commercially available temperature-controlled microscope stage. Using this adaptor, we observed temperature-dependent organelle dynamics in living plant cells; morphological changes in chloroplasts and peroxisomes were temperature dependent. This newly developed adaptor can be easily placed on a temperature-controlled stage to capture intracellular responses to temperature at unprecedentedly high resolution.