First-in-Human Study for Evaluating the Accuracy of Smart Ring Based Cuffless Blood Pressure Measurement
Jihoon Kim1, Sung-A Chang1, Seung Woo Park2
1Division of Cardiology, Heart Vascular Stroke Institute, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.
Background:
Recently, a ring-type cuffless blood pressure (BP) measuring device has been developed. This study was a prospective, single arm, first-in-human pivotal trial to evaluate accuracy of BP measurement by the new device.
Methods:
The ring-type smart wearable monitoring device measures photoplethysmography signals from the proximal phalanx and transmits the data wirelessly to a connected smartphone. For the BP comparison, a cuff was worn on the arm to check the reference BP by auscultatory method, while the test device was worn on the finger of the opposite arm to measure BP simultaneously. Measurements were repeated for up to three sets each on the left and right arms. The primary outcome measure was mean difference and standard deviation of BP differences between the test device and the reference readings.
Results:
We obtained 526 sets of systolic BP (SBP) and 513 sets of diastolic BP (DBP) from 89 subjects, with ranges of 80 to 175 mmHg and 43 to 122 mmHg for SBP and DBP, respectively. In sample-wise comparison, the mean difference between the test device and the reference was 0.16 ± 5.90 mmHg (95% limits of agreement [LOA], -11.41, 11.72) in SBP and -0.07 ± 4.68 (95% LOA, -9.26, 9.10) in DBP. The test device showed a strong correlation with the reference for SBP (r = 0.94, P < 0.001) and DBP (r = 0.95, P < 0.001). There were consistent results in subject-wise comparison.
Conclusion:
The new ring-type BP measuring device showed a good correlation for SBP and DBP with minimal bias compared with an auscultatory method.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: