Comparison of cuff inflation and cuff deflation brachial sphygmomanometry with intra-arterial blood pressure as
Branko G Celler1, Andy Yong2,3, Imants Rubenis2
1Biomedical Systems Research Laboratory, University of New South Wales.
Abstract:
Conventional sphygmomanometry with cuff deflation is used to calibrate all noninvasive BP (NIBP) instruments and the International Standard makes no mention of calibrating methods specifically for NIBP instruments, which estimate systolic and diastolic pressure during cuff inflation rather than cuff deflation. There is however increasing interest in inflation-based NIBP (iNIBP) instruments on the basis of shorter measurement time, reduction in maximal inflation pressure and improvement in patient comfort and outcomes. However, we have previously demonstrated that SBP estimates based on the occurrence of the first K1 Korotkoff sounds during cuff deflation can underestimate intra-arterial SBP (IA-SBP) by an average of 14 ± 10 mmHg. In this study, we compare the dynamics of intra-arterial blood pressure (IABP) measurements with sequential measurement of Korotkoff sounds during both cuff inflation and cuff deflation in the same individual. In 40 individuals aged 64.1 ± 9.6 years (range 36-86 years), the overall dynamic responses below the cuff were similar, but the underestimation error was significantly larger during inflation than deflation, increasing from 14 ± 10 to 19 ± 12 mmHg ( P < 0.0001). No statistical models were found which could compensate for this error as were found for cuff deflation. The statistically significant BP differences between inflation and deflation protocols reported in this study suggest different behaviour of the arterial and venous vasculature between arterial opening and closing which warrant further investigation, particularly for iNIBP devices reporting estimates during cuff inflation. In addition, measuring Korotkoff sounds during cuff inflation represents significant technical difficulties because of increasing pump motor noise.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.