Electrode Montage for Bilateral Cervical Vestibular-Evoked Myogenic Potential Testing
Jessie N Patterson1, Nour El Hidek2, Kristen L Janky1
1Department of Audiology, Boys Town National Research Hospital, Omaha, Nebraska.
Background:
Cervical vestibular-evoked myogenic potentials (cVEMPs) are predominantly ipsilateral, myogenic responses originating from saccular activation. Some individuals have contralateral-crossed cVEMP responses with monaural air-conducted stimulation (ACS) which can contaminate cVEMP responses with bilateral stimulation. While the origin of the contralateral-crossed response is under debate, its presence has implications for cVEMP testing with midline bone conduction vibration (BCV).
Purpose:
The purpose of this study was to determine the origin of the contralateral-crossed cVEMP response. It was hypothesized that the crossed response is due to electrode contamination and would disappear with a modified electrode montage.
Research Design:
This is a cross-sectional research study.
Study Sample:
Fifteen healthy participants (30 ears; mean age: 27.4 19-39; 10 females).
Data Collection And Analysis:
Participants completed cVEMP testing using three stimulation methods (monoaural ACS, binaural ACS, and midline BCV) and two electrode montages (sternum reference and Fp reference).
Results:
In the monoaural ACS with sternum reference condition, 53.3% ears had contralateral-crossed cVEMP responses that were in-phase with the ipsilateral response for all but three ears. Whereas in the monoaural ACS with Fp reference condition, 3% had a contralateral-crossed cVEMP response. ACS and BCV cVEMP corrected amplitudes were significantly larger in the sternum reference conditions, which is attributed to artificial enhancement from the in-phase contralateral-crossed responses.
Conclusions:
The significant reduction of contralateral-crossed responses in the Fp reference condition suggests that the contralateral-crossed cVEMP response is due to reference electrode contamination and may be a more appropriate reference placement when completing cVEMPs with midline BCV.