Size characterization of x-ray tube source with sphere encoded imaging method
Jian Yu1, Dikai Li1, Yanmeng Dai1
1Shenzhen Key Laboratory of Ultraintense Laser and Advanced Material Technology, Center for Advanced Material Diagnostic Technology, and College of Engineering Physics, Shenzhen Technology University, Shenzhen 518118, China.
Abstract:
In x-ray imaging, the size of the x-ray tube light source significantly impacts image quality. However, existing methods for characterizing the size of the x-ray tube light source do not meet measurement requirements due to limitations in processing accuracy and mechanical precision. In this study, we introduce a novel method for accurately characterizing the size of the x-ray tube light source using spherical encoded imaging technology. This method effectively mitigates blurring caused by system tilting, making system alignment and assembly more manageable. We employ the Richardson-Lucy algorithm to iteratively deconvolve the image and recover spatial information about the x-ray tube source. Unlike traditional coded imaging methods, spherical coded imaging employs high-Z material spheres as coding elements, replacing the coded holes used in traditional approaches. This innovation effectively mitigates blurring caused by system tilting, making system alignment and assembly more manageable. In addition, the mean square error is reduced to 0.008. Our results demonstrate that spherical encoded imaging technology accurately characterizes the size of the x-ray tube light source. This method holds significant promise for enhancing image quality in x-ray imaging.