Uncommon causes of hemoglobin E flags identified during measurement of hemoglobin A1c by ion-exchange
Thomas Herb1, Alexander S Taylor1, Shih-Hon Li1
1Department of Pathology, University of Michigan, Ann Arbor, MI, US.
We present 3 cases of discordant results from screening hemoglobin A1c (HbA1c) measured by ion-exchange high-performance liquid chromatography (HPLC) all due to various forms of interference and flagged by the instrument as "suspected hemoglobin E (HbE)." The first case was due to a rare hemoglobin variant, later confirmed to be hemoglobin Hoshida, the second due to "true" heterozygous HbE, and the third a result of analytical artifact causing splitting of the HbA1c peak without an underlying variant hemoglobin. We examine the similarities in these cases along with the laboratory work-up to classify each cause of interference to demonstrate the wide array of potential causes for the suspected HbE flag and why it warrants proper work-up. Because there is no standardized method of reporting out hemoglobin variant interference in HbA1c measurement, we discuss our laboratory's process of investigating discordant HbA1c measurements and reporting results in cases with variant interference as 1 possible model to follow, along with discussing the associated laboratory, ethical, and clinical considerations. We also examine the structure of hemoglobin Hoshida, HbE, and conduct a brief literature review of previous reports.
We present 3 cases of discordant results from screening hemoglobin A1c (HbA1c) measured by ion-exchange high-performance liquid chromatography (HPLC) all due to various forms of interference and flagged by the instrument as "suspected hemoglobin E (HbE)." The first case was due to a rare hemoglobin variant, later confirmed to be hemoglobin Hoshida, the second due to "true" heterozygous HbE, and the third a result of analytical artifact causing splitting of the HbA1c peak without an underlying variant hemoglobin. We examine the similarities in these cases along with the laboratory work-up to classify each cause of interference to demonstrate the wide array of potential causes for the suspected HbE flag and why it warrants proper work-up. Because there is no standardized method of reporting out hemoglobin variant interference in HbA1c measurement, we discuss our laboratory's process of investigating discordant HbA1c measurements and reporting results in cases with variant interference as 1 possible model to follow, along with discussing the associated laboratory, ethical, and clinical considerations. We also examine the structure of hemoglobin Hoshida, HbE, and conduct a brief literature review of previous reports.
Related Concept Videos
Ion-Exchange Chromatography
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Hemoglobin
When all four heme groups are bound to oxygen, the resulting molecule is called oxyhemoglobin. As a result, arterial blood...
Disorders of Erythrocytes
Erythrocyte disorders can be broadly categorized into two main types: anemic and polycythemic conditions.
A low oxygen-carrying capacity of the blood due to the loss, lower production, or destruction of erythrocytes is termed anemia. Hemorrhagic anemia, for example, occurs when bleeding from an external wound or internal ulcer reduces erythrocyte counts.
On the other...