Related Experiment Video
Updated: Jun 20, 2026

A Novel Method: Super-selective Adrenal Venous Sampling
Published on: September 15, 2017
Validating the usefulness of Sudoscan in ATTRv: a single centre experience
Elena Moreno-Moraleda1, Juan González-Moreno2,3,4, Eugenia Cisneros-Barroso5
1Universitat de Les Illes Balears, Palma, Spain.
Background:
Variant transthyretin amyloidosis (ATTRv) can cause sensorimotor and autonomic neuropathy. Objective quantification of sudomotor function may be essential for early diagnosis and early initiation of treatment. The aim of this study is to evaluate the diagnostic value of the Sudoscan® in ATTRv.
Methods:
Electrochemical skin conductance (ESC) was measured in V30M ATTRv patients, asymtomatic V30M carriers and healthy controls. Comparisons between the three groups were made using the Kruskal-Wallis test, and ROC curves were used to estimate the discriminatory power of ESC values between groups.
Results:
ESC was measured in 52 ATTRv patients, 107 asymptomatic carriers and 40 healthy controls. ESC was significantly lower in ATTRv patients compared to asymptomatic carriers and healthy controls in both feet and hands; median values are as follows: 40 µS, 78 µS and 81 µS, respectively (p < 0.001), and 53 µS, 69 µS and 74 µS, respectively (p < 0.001). ESC in feet < 70.5 µS had a sensitivity of 89.7% and specificity of 84.6% to discriminate asymptomatic carriers from patients with ATTRv.
Conclusion:
The determination of ESC by Sudoscan® is a rapid, noninvasive and easily reproducible technique capable of discriminating patients with ATTRv from asymptomatic carriers and healthy controls with adequate sensitivity and specificity.
More Related Videos
09:07Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
05:04Author Spotlight: Evaluating Clinicians' Adoption of Ultrasound-Guided Vascular Cannulation Through Simulation Training
Published on: August 9, 2024
Related Concept Videos
Pulse Assessment Sites
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...