Accuracy of tidal volume delivery by paediatric intensive care ventilators: A bench-model study
Meryl Vedrenne-Cloquet1, Samuel Tuffet2, Bruno Louis3
1Pediatric Sleep and Noninvasive Ventilation Unit, EA 7330 VIFASOM, AP-HP, Hôpital Necker Enfants-Malades, Paris, France; Université de Paris, Paris, France; Réanimation Médicochirurgicale Pédiatrique, AP-HP, CHU Necker-Enfants Malades, Paris, France.
Background:
Tidal volume (Vt) delivery during mechanical ventilation is influenced by gas compression, humidity, and temperature.
Objectives:
This bench study aimed at assessing the accuracy of Vt delivery by paediatric intensive care ventilators according to the humidification system. Secondary objectives were to assess the following: (i) the accuracy of Vt delivery in ventilators with an integrated Y-piece pneumotachograph and (ii) the ability of ventilators to deliver and maintain a preset positive end-expiratory pressure.
Methods:
Six latest-generation intensive care ventilators equipped with a paediatric mode were tested on the ASL5000 test lung in four simulated paediatric bench models (full-term neonate, infant, preschool-age chile, and school-age child), under volume-controlled mode with a heated humidifier (HH) or a heat moisture exchanger, with various loading conditions. Three ventilators equipped with a Y-piece pneumotachograph were tested with or without the pneumotachograph in the neonatal and infant models. "Accurate Vt" delivery was defined as a volume error (percentage of the preset Vt under body temperature and pressure and saturated water vapour conditions) being ≤10 % of the absolute preset value.
Results:
Vt accuracy varied significantly across ventilators but was acceptable in almost all the ventilators and all the models, except the neonatal model. The humidification system had an impact on Vt delivery in the majority of the tested conditions (p < 0.05). The use of an HH was associated with a better Vt accuracy in four ventilators (V500, V800, R860, and ServoU) and allowed to achieve an acceptable level of volume error in the neonatal model as compared to the use of heat moisture exchanger. The use of an integrated pneumotachograph was associated with lower volume error in only one ventilator (p < 0.01). All the tested ventilators were able to maintain adequate positive end-expiratory pressure levels.
Conclusion:
The humidification system affects Vt accuracy of paediatric intensive care ventilators, especially in the youngest patients for whom the HH should be preferred.
More Related Videos
06:22Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome ARDS
Published on: April 7, 2021
05:56Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...