Polarity-Dependent, Reduced or No-Output Shocks in Implantable Cardioverter-Defibrillators
Charles D Swerdlow1, Michael Ringle2
1Smidt Heart Center, Cedars Sinai Medical Center, Los Angeles, California, USA.
Abstract:
Recently, polarity-dependent shock failures were reported in implantable cardioverter-defibrillators caused by structural failure in the high-voltage feedthrough. Short circuits may occur when the right ventricular coil is cathodal for phase 1 of biphasic shocks (cathodal shock). This viewpoint proposes a mechanism for observed polarity dependence and considers whether the same mechanism may apply in other shock-induced, short circuits. Implantable cardioverter-defibrillator connections to the lead traverse feedthroughs into the hermetically sealed housing ("Can"). The feedthrough comprises 2 concentric, conducting metal cylinders, the inner pin-conductor to the right ventricular coil and outer Can, separated by impermeable insulation. Shock failure depends on 3 conditions: 1) development of a fluid layer in the feedthrough, creating a conduction path in parallel with the shock pathway; 2) the radial gradient of the electric field in the fluid, so resistive heating during a shock vaporizes water to form a high-resistance gas bubble around the pin; and 3) field emission of electrons at the cathode, with rate and energy dependent on the field's strength and the cathode's potential-energy barrier to emission. For cathodal shocks, electrons emitted at the metal pin may initiate an ionization avalanche in the gas until it "breaks down" into a low-resistance plasma, resulting in a short circuit. For anodal shocks, the effective cathode is the liquid-gas interface, where the field is weaker than at the pin. Additionally, solvated electrons in aqueous solution must overcome a higher potential-energy barrier to be emitted. This permits the high-resistance gas bubble to stabilize so that the shock is completed.
More Related Videos
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Electrophysiology of Normal Cardiac Rhythm
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
Antiarrhythmic Drugs: Class II Agents as β-Adrenergic Blockers
Clipper Circuit
The operation of a clipper circuit can be exemplified by analyzing a dual-clipper configuration setup that integrates two ideal diodes, each paired with a biasing...
Node Analysis for AC Circuits
To unravel the complexities of this system, nodal analysis is employed, a powerful technique founded on Kirchhoff's current law (KCL), which remains valid for phasors. AC circuits can effectively be...