A functional brace for treating Colles fractures; volunteer trial of a new design
Douglas Wardlaw1, Ann H Murdoch2
1Robert Gordon University, Aberdeen, United Kingdom; NHS Grampian, Address, 109 Inchgarth Road, Pitfodels, Cults, Aberdeen AB15 9NX, United Kingdom.
Background:
Presently used Colles' fracture treatments have similar outcomes with significant complications. Previous studies of a dynamic functional fracture brace, achieving similar or better results had no significant complications. A novel brace design is described to achieve optimal patient outcomes.
Methods:
Patient forearm measurements combined with data from a previously tested brace provided parameters for a brace formed using computer assisted design. Fracture swelling was simulated using a 40 ml (ml) plastic water filled bag placed over the dorsum of the lower radius. The interface pressure between the brace, skin and bag were measured on twelve human volunteers, four with forearm measurements to fit each of small, medium and large brace sizes. Measurements were taken at intervals reducing volume of fluid in the bag by 10 ml until empty, first wearing a brace, and then with a Colles plaster of Paris cast.
Findings:
The brace produced interface pressures over the lower radius in the range of 62.5-90 (mm Hg) and when the bag was completely empty the pressure range was 43-83 mmHg. The initial interface pressures in the Colles casts were in the range of 15-18 mm of Mercury (mm Hg) and when 10 mls were removed from the plastic bag the interface pressure dropped immediately to almost zero.
Interpretation:
The brace measurements show pressures higher than a Colles cast which compensate for reduction of simulated swelling. It is safe and is potentially the optimal design to achieve significantly improved patient outcomes.
More Related Videos
06:59Fracture Apparatus Design and Protocol Optimization for Closed-stabilized Fractures in Rodents
Published on: August 14, 2018
07:35Creating Rigidly Stabilized Fractures for Assessing Intramembranous Ossification, Distraction Osteogenesis, or Healing of Critical Sized Defects
Published on: April 11, 2012
Related Concept Videos
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Flail Chest-II
Assessment:
1. Clinical Evaluation:
History: